![]() ![]() The item will arrive secured to a pallet and delivered up to your address. Freight: For oversized items, the freight cost is $545 per item.Alarm Disabled During Priming: Air-in line.Maximum Volume Infused Under Single-Fault Conditions: CADD administration set - 0.15 ml.Actual Test Data for Bolus Accuracy at 20 ml:.Actual Test Data for Bolus Accuracy at 0.05 ml:.Accumulated Air: Greater than 1 ml air over 15 minutes (nominal).High - Single bubble greater than 150 μL.Low - Single bubble greater than 400 μL.High Pressure Alarm Threshold: 18 ± 9 psi (1.24 ± 0.62 bar).During the total infusion time, the accuracy averages out. ![]() At low infusion rates, this accuracy may not be achieved for short periods. System Delivery Accuracy: ± 6% (nominal).Atmospheric Pressure: 70 to 106 kPa (10.2 to 15.4 psi).Relative Humidity: 20 to 90% relative humidity, non-condensing.System Storage and Transportation Temperature: -20 to 60 ☌ (-4 to 140 ☏).System Operating Temperature: 2 to 40 ☌ (36 to 104 ☏).Power Sources: 4 AA (IEC LR6) alkaline batteries, AC adapter, rechargeable battery pack.Note: Values are nominal and based on actual test data.10 ml/hr - medication cassette reservoir: 44 seconds.0.1 ml/hr - medication cassette reservoir: 108 minutes.Maximum Infusion Pressure: 27.0 psi (1.86 bar).Moisture Protection: Splashproof (IPX4) per IEC 60529. ![]() This is only enabled while the pump is in run mode.
0 Comments